Comparison of new point-of-care troponin assay with high sensitivity troponin in diagnosing myocardial infarction.

Link to article at PubMed

Related Articles

Comparison of new point-of-care troponin assay with high sensitivity troponin in diagnosing myocardial infarction.

Int J Cardiol. 2014 Nov 15;177(1):182-6

Authors: Aldous S, Mark Richards A, George PM, Cullen L, Parsonage WA, Flaws D, Florkowski CM, Troughton RW, O'Sullivan JW, Reid CM, Bannister L, Than M

Abstract
OBJECTIVES: The aim of this study is to compare a new improved point of care cardiac troponin assay (new POC-cTnI) with 1. its predecessor (old POC-cTnI) and 2. a high sensitivity assay (hs-cTnI) for the diagnosis of acute myocardial infarction (AMI) and for major adverse cardiac events (MACE) by 30 days.
METHODS: This is a single centre observational study, set in Christchurch Hospital, New Zealand. Patients presenting to the emergency department with non-traumatic chest pain underwent blood sampling at 0 h and 2h post presentation for analysis with the 3 cTnI assays for the outcome of AMI and for analysis using an accelerated diagnostic protocol (ADP-normal 2h troponins, normal electrocardiograms and Thrombolysis In Myocardial Infarction (TIMI) score of 0 or ≤ 1) for 30 day MACE.
RESULTS: Of 962 patients, 220 (22.9%) had AMI. Old POC-cTnI was least sensitive at 70.0% (65.4-73.9%) by 2h (p<0.001). New POC-cTnI, sensitivity 93.6% (89.9-96.2%) had similar sensitivity to hs-cTnI, sensitivity 95.0% (91.5-97.3%) (p = 0.508). There were 231 (24.0%) patients with 30 day MACE. When used as part of the ADP, all assays had 100% (98.0-100%) sensitivity using TIMI = 0. Sensitivities of new POC-cTnI ADP, 98.3% (95.4-99.4%), old POC-cTnI, 96.5% (93.2-98.4%) and hs-cTnI, 98.7% (96.0-99.7%) were similar (p = 0.063-0.375) using TIMI ≤ 1.
CONCLUSIONS: A new POC-cTnI has improved sensitivity for AMI and MACE compared with its predecessor and comparable sensitivity to a high sensitivity assay. Now that sensitivities of the POC assay are improved, the new assay may be a useful alternative to central laboratory assays when rapid turn-around times are not possible.

PMID: 25499373 [PubMed - indexed for MEDLINE]